Mode; Cells / Μl; Min And Max; Select Numeric Precision - Beckman Coulter Navios Instructions For Use Manual

Flow cytometer
Table of Contents

Advertisement

Using Navios Software
Analysis Menu

Mode

The mode channel is the channel with the largest number of counts between region markers.
Gate colors can be customized using the drop down color box and selecting the required color.
Cells / μL
This displays the absolute count of a population based on the calibration factor (CAL Factor) and the
number of particles within a user defined CAL region.

Min and Max

Displays linear region boundaries.

Select Numeric Precision

Choose the precision you want to specify for, Percent, Mean or CV.

Enter Numeric Value

Maximum 3 decimal places for CV, 4 decimal places for Percent and maximum 5 digits for Mean.

Advanced Statistics Configuration

Selecting the Advanced button offers further options to
select the way that the mean is calculated.

Log-Log

Log-Lin
Automatic Statistic Percentiles

Differences Between Log-Log and Log-Lin Statistical Calculations

For most practical applications users may never notice the difference between the two. In general,
Log-Log mean appears to have its value skewed further to the right than a Log-Lin mean calculation
on the same data. This skew is more pronounced when the distribution of data extends greater than
±half a decade from the mean value, however calculated.
Log-Log
Selecting Log-Log calculates the mean values of Log data using geometric conversion of each
individual channel before a mean value is calculated.
10-142