Display Range; Limit Values; Language; Date / Time - Proceq DIGI-SCHMIDT 2000 ND Operating Instructions Manual

Concrete test hammer
Hide thumbs Also See for DIGI-SCHMIDT 2000 ND:
Table of Contents

Advertisement

Mx: Delete the Lowest and Highest Values
Only possible from EPROM version 2.0. The amount "m"
of both the lowest and highest values are deleted.
• Under "m", enter the number of measured values to be
eliminated. See also the number of test impacts "n"
set, under "Mean Value", page 6.
Example: If "m" = 3, then at least 12 should be set for "n".
4.6

Display Range

The display range for the rebound value R on the display
can be adapted to the testing range.
• Enter R
and R
.
min
max
4.7

Limit Values

The limit marks on the display serve to aid fast recogni-
tion of deviations from the measured values.
The range within which the rebound values R are to lie
can be selected (see Fig. 3.1 "Measuring display",
page 5).
The selected limit values are not saved!
• Enter R
and R
.
min
max
© 2001 PROCEQ SA
4.8

Language

• The display texts can be selected in the language
required.
4.9

Date / Time

The inbuilt clock saves the measured values together
with the test date and time.
Correct the values set, if necessary.
4.10

Data Output

See under "Data Output", page 11.
When all the settings have been made on the display
unit, press the END key. The device is ready to continue
with the "Measuring Process", page 10.