SECTION 7—BASE COMPONENTS
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 25.
Wire
Jumper
Wires
Wire and Jumper
Connectors

Testing/Replacing Capacitor

Testing

The capacitor is an electrical storage device that stores an electronic charge.
ALWAYS discharge the capacitor before testing or handling. Failure to discharge
the capacitor may result in bodily harm and/or damage to test equipment.
NOTE: For this procedure, refer to FIGURE 7.6 on page 39.
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
2. Remove the two wire connectors from the capacitor.
3. Discharge the capacitor by placing a screw driver with an insulated handle across the
wire connectors on the top of the capacitor.
4. Connect the leads of a capacitance meter to the terminals on the capacitor.
NOTE: For the capacitor to be within specification, the meter should read 3 mfd +10% or –5%.
5. Perform one of the following:
• If the capacitor is out of specification, replace the capacitor. Refer to Replacing
Capacitor on page 39.
• If the capacitor is in specification, reconnect the two wires and proceed to STEP 7.
6. Install the cabinet. Refer to Removing/Installing the Cabinet on page 25.
®
HomeFill
II AW Compressor
Base
FIGURE 7.5 Replacing Transformer
WARNING
38
Transformer
Mounting Screw
Part No 1154316