Invacare HomeFillII AW Service Manual page 62

Table of Contents

Advertisement

SECTION 10—COMPRESSOR REPLACEMENT
12. Using diagonal cutters, cut the tie wrap that secures the inlet hose to the inlet of
stage 1 (FIGURE 10.4).
13. Disconnect the oxygen inlet hose from the inlet of stage 1 (FIGURE 10.4).
14. Using diagonal cutters, cut the tie wrap that secures the oxygen inlet hose to the
receptacle fitting (FIGURE 10.4).
15. Remove the oxygen inlet hose from the receptacle fitting (FIGURE 10.4).
16. Using diagonal cutters, cut the tie wrap that secures the oxygen sensor tubing to the
oxygen inlet hose (FIGURE 10.4).
17. Remove the oxygen sensor tubing from the oxygen inlet hose (FIGURE 10.4).
18. Note where each of the four wires and two jumpers are connected to the transformer.
19. Using needle nose pliers, disconnect the connectors from the transformer. See Detail
"A" in FIGURE 10.4.
20. Using needle nose pliers, disconnect the two connectors from the power inlet
(FIGURE 10.4).
21. Using needle nose pliers, disconnect the two connectors (not shown) from the hour
meter (FIGURE 10.4).
22. Lift control panel assembly off of base and set aside (FIGURE 10.4).
®
HomeFill
II AW Compressor
62
Part No 1154316