SECTION 9—MOTOR REPLACEMENT
SECTION 9—MOTOR
REPLACEMENT
Tools Required:
7
/
open ended wrench,
16
Phillips screwdriver

Disassembly

NOTE: For this procedure, refer to FIGURE 9.1 through FIGURE 9.6.
1. Install an empty cylinder onto the compressor to release the pressure in the lines.
2. Remove the cylinder. Refer to Removing/Installing the Cabinet on page 25.
3. Disconnect power from the HomeFill unit.
4. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
5. Disconnect the two pin connector that connects the motor to the wiring harness
(FIGURE 9.1).
6. Using needle nose pliers, disconnect the two wires from the capacitor (FIGURE 9.1).
7. Using needle nose pliers, disconnect the grounding clip from the motor (FIGURE 9.1).
Control Panel
Two Pin
Gusset
Connector
(STEP 4)
Motor
Grounding
Clip
Wires (STEP 5)
®
HomeFill
II AW Compressor
5
/
open ended wrench, Diagonal Cutters, Needle Nose Pliers
8
Capacitor
Mounting Screw
(Hidden behind wiring assembly)
FIGURE 9.1 Motor Replacement
52
Control Panel
Gusset
Mounting
Screw
Part No 1154316