SECTION 4—COMPRESSOR OPERATION

Disconnecting Cylinder From Compressor

DO NOT modify any connections on the HomeFill compressor. NEVER use tools of
any kind to connect/disconnect the cylinder and the compressor. Otherwise, severe
injury and/or damage may occur.
NOTE: For this procedure, refer to FIGURE 4.3 on page 22.
1. Press the compressor power switch to the Off (O) position.
2. Grasp the cylinder/regulator assembly in the area behind the cylinder gauge.
3. With the other hand, grasp the outer ring (sleeve) of the connector fillport and push
down.
4. Lift up on the cylinder/regulator assembly to remove from the connector fillport.
5. When the cylinder fillport is disconnected from the connector fillport, release the
outer ring (sleeve) of the connector fillport and use two hands to remove the cylinder
from the compressor cradle.
- The fillport cover on the connector fillport and the cylinder fillport MUST be
replaced after filling and whenever not in use. If either of the fillport covers are
missing replace these parts before using or returning to the user.
6. Place the fillport covers onto the connector and cylinder fillports.
NOTE: The fillport covers should be used whenever cylinders are not being filled.
Cylinder Gauge
Compressor
Cradle
Connector Fillport
FIGURE 4.3 Disconnecting Cylinder From Compressor
®
HomeFill
II AW Compressor
WARNING
WARNING
Cylinder/Regulator
Assembly
Cylinder
Fillport
Power Switch
22
PUSH DOWN
Outer Ring
(Sleeve)
Part No 1154316