will proceed to 14 days and then to 21 days
and then to 30 days before it shows OFF
again. The maximum length of time between
scheduled bump tests is 30 days.
To disable the bump due warning, set the
interval to OFF.
Once the bump due interval
has been set correctly, click
the MODE button three times
in rapid succession to return to
the Main Menu.
Note: The Bump Test Interval and Bump
Test Reminder are used exclusively with
the MultiPro IQ Express Dock.
6.9

Setup accept

Once the settings in any of the submenus
have been changed as needed, return to the
Main Menu by clicking the MODE button three
times in rapid succession. Once in the Main
Menu hold the MODE button
down for three seconds to exit.
The MultiPro will provide a
countdown to accept changes
to the settings.
Click MODE to accept the
changes and the instrument will
show "setup stored" and
proceed to turn itself off. Allow
the countdown to run out if you
do not wish to save the
changes.