Function Check; What Do I Do If - Seca 727 Instruction Manual And Guarantee

Wireless baby scale
Hide thumbs Also See for 727:
Table of Contents

Advertisement

9. FUNCTION CHECK

10.WHAT DO I DO IF ...?

Fault
... no weight is dis-
played when scale is
loaded?
... 0.000 does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
74 •
Perform a function check prior to each use.
A complete function check includes:
• visual inspection for mechanical damage
• checking the alignment of the device
• visual and function check of the display elements
• function check of all the controls shown in the
section entitled "Overview"
• function check of optional accessories
If you notice any faults or deviations during the function
check, first try to resolve the error with the aid of the
section entitled "What do I do if ..." in this document.
CAUTION!
Personal injury
If you notice any faults or deviations during the
function check which cannot be resolved with
the aid of the section entitled "What do I do if
..." in this document, you may not use the
device.
– Have the device repaired by seca service or
by an authorized service partner.
– Follow the section entitled "Servicing" in this
document.
Cause/Remedy
The scale is not connected to the power supply.
- Check if the scale is switched on.
- Check whether the battery block is inserted
and that the plug contacts are properly con-
nected.
A load was placed on the scale before it was
switched on.
- Remove load.
- Switch scale off and back on again.
The relevant segment is faulty.
- Please notify maintenance service.
Battery block power is running down.
- Charge battery block.

Hide quick links: