This method is recommended to keep the equipment clean and free of contaminants when used between patients for
treatment of non-intact skin or incontinence management, where there is an increased risk of patient cross-
contamination.
After each use, clean common contact surfaces, such as control panel and ultrasound transducer, with ACP
germicidal wipes.
With a second ACP germicidal cloth, wipe again leaving surfaces wet for at least 5 minutes. Allow the
surface to air dry before patient use.
Barriers should also be used on the equipment for treatment of non-intact skin or incontinence
management. This technique will inactivate M. tuberculosis as well as most bacteria and viruses.