Testing DC Current

Use the Wavetek 1281 and the precision shunt to measure the 525A output as
shown in Figure 6-2 . Take the Voltage reading from the Wavetek 1281 and divide
it by the 742A-1 actual value.
525 A
TEMPERATURE / PRESSURE CALIBRATOR
525A
mA
VOLTS
RTD
100mA MAX
HI
HI
OUTPUT
100V MAX
LO
LO
20V PK
MAX
4W RTD
HI
INPUT
SENSE
LO
20V PK
MAX
mA
Output
Terminals
100 mA output
(A)
0
0.025
0.075
0.1
6-6
STBY
VOLTS
TC
TYPE
OPR
mA
RTD
UNITS
TC
INPUT/OUTPUT
OUTPUT
INPUT
ZERO
7
8
9
SHIFT
SETUP
CJC
C / F
4
5
6
SET
RECALL
AUTOSET
20V PK
MAX
1
2
3
RNG LOCK
LOCAL
EXP
/
0
ENTER
Figure 6-2. Measuring DC Current
Table 6-3. Measuring DC Current
Volt
Current Shunt
CE
Curent
742A-1
(I=E/R)
Shunt Value
Direct into 1281
current input
1281 DCV Function
1 Year (A)
0.000002
4.5E-06
9.5E-06
0.000012
ajr25f.eps