Remote With Lockout State - Fluke 525A User Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525A:
Table of Contents

Advertisement

Front panel operation is disabled except for the LOCAL (0 key). Pressing 0, using
RS-232 to send the command LOCAL, or IEEE-488 to send the GTL (Go To
Local) message returns the Calibrator to the local state.

Remote with Lockout State

When the Calibrator is placed in lockout, either via RS-232 LOCKOUT command,
or via the IEEE-488 message LLO, the 525A front panel controls are totally
locked out. The top line of the display changes to: REM.
To return the Calibrator to the local with lockout state, send the RS-232 LOCAL
command or the IEEE-488 GTL (Go To Local) message.
Table 4-2 summarizes the possible Operating state transitions. (For more
information on IEEE-488 GPIB messages, see "IEEE-488 Overview."
From
Local
Remote
Local with
Lockout
Remote with
Lockout
Changing Between Remote and Local Operation
Table 4-2. Operating State Transitions
To
Remote
Local with
Lockout
Local
Local softkey
Remote with
Lockout
Local
Remote with
Lockout
Local
Local with
Lockout
Front Panel
MLA (REN True)
LLO
GTL or REN
False
LLO
REN False
MLA (REN True)
REN False
GTL
Remote Operation
GPIB
Serial
Message
Command
REMOTE
LOCKOUT
LOCAL
LOCKOUT
LOCAL
REMOTE
LOCAL
4
4-9