The 1732 Oxygen Transmitter has 4 alarm relays, a built in alarm annunciator and an alarm log. When an
alarm occurs, the Alarm Light will flash. To find out what the alarm is, press the A
∆ button has been pressed, the transmitter goes into the Alarm Display Mode. In this mode
When the A
LARM
some of the buttons take on a different function.
Button text
S
/ R
ETUP
UN
D
/ F
ISPLAY
UNCTION
D
/ F
ISPLAY
UNCTION
A
/ O
LARM
PTION
A
/ O
LARM
PTION
A
A
/ E
LARM
CCEPT
NTER
G
1 P
1 / S
AS
URGE
ENS
G
2 P
2 / A
AS
URGE
UTO
* This button is not used in the Alarm Display Mode
When the Alarm Mode has been entered, the Setup Light flashes once a second. The transmitter will return
to Run Mode if the S
ETUP
All relays have fail-safe alarm contacts. That is –
When the transmitter is powered off the contacts are open circuit
When the transmitter is powered on but there are no alarms the contacts will be closed
When there is a current unaccepted alarm event the contacts will be open circuit
When there is a current accepted alarm event the state of the contacts will depend on the selection
in the Commissioning Menu #53. Refer to the 1732 Technical Manual for details.
All alarms drive the alarm light on the front door.
The light will be off if there are no alarms current
The light will flash if there is a current alarm that has not been acknowledged
The light will be on steady if there are current alarm(s) that have not been cleared
The light will flash faster as more alarms occur
February 2015
1732 Oxygen Transmitter
Run Mode
*
*
*
Enter Alarm Display Mode
Enter Alarm Log Mode
*
I
*
MP
C
*
AL
/ R
button is pressed a second time, or after 60 seconds of keypad inactivity.
UN
∆ button.
LARM
Alarm Mode
Return to Run Mode
Next alarm
Last alarm
Alarm activated time
Alarm acknowledged time
Acknowledge alarm
*
*
6. A
LARMS
Operators Manual
19