6.3 Weighing Units

Weighing results can be displayed in different units.
Change by pressing the
Press and hold the
change after a few seconds.
Grams *
Ounce
Troy ounce
Pennyweight

6.4 Zero-Tracking

The Auto-Zero function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be
weighed, incorrect weighing results can be displayed due to the "stability
compensation". (Example: Slowly draining fluids from a container on the balance).
When apportioning involves small variations of weight, it is advisable to switch off this
function.
If Zero-Tracking however is switched off, the weighing display becomes more busy.
Activate/deactivate Zero-Tracking:
 When the balance is switched off, press the OFF-button and keep it pressed
 Do not release OFF -button. Press also the ON/TARE-button and keep it pressed
 Release OFF –button briefly and press again
 "AF" will appear on the screen
 Release both buttons. The balance is situated in the menu.
 Select "tr" with the help of the ON/TARE-button
 Confirm by pressing the OFF-button. The current setting will be shown on the
screen.
EMB-S-BA-e-1612
button.
button when the weighing plate is loaded and the unit will
Display
Conversion factor
screen
1 g =
g
1.
oz
0.035273962
ozt
0.032150747
dwt
0.643014931
13