1587 FC/1587/1577
Calibration Manual
8. Secure the PCA to the LCD assembly by installing the two screws that attach
Reassembling the Meter
To reassemble the Meter, logically reverse the previous disassembly procedures.
In the process, make sure to re-establish all electrical connections; specifically,
the elastomeric strip for the LCD, the red and black plastic shields for the input
terminals, and the two spring contacts for the Insulation Test Probe. Also make
sure all parts are correctly aligned and positioned; do not force-fit any of the parts
into position.
20
panel. After the LCD assembly is in position, lock it in place by pressing
(below the display) the bezel against the PCA; listen for the tab on the bezel
to snap (lock) into position on the PCA.
the PCA to the bezel. These screws also ensure contact between the glass
LCD display, the elastomeric strip, and the PCA.
Figure 6. Accessing the LCD
Screws
Gray Plastic
Tab
PCA
Light-dispersing
Back Panel
Glass LCD
Display
Elastomeric
Strip
Plastic
Bezel
ebp05f.eps