Tektronix TDS5000B Series Read This First Manual page 103

Digital phosphor oscilloscope
Hide thumbs Also See for TDS5000B Series:
Table of Contents

Advertisement

Commands
DIAg:CONTROL:HALT
Related Commands
Syntax
Arguments
Example
5- 20
Table 5- 5: DIAgnostic Commands
Command
DIAg:STOP
TEST
TEST:RESults?
TEST:RESults:VERBose?
TEST:STOP
This command (no query form) determines whether the next execution of
diagnostics will stop on the first diagnostic failure that occurs or will execute the
selected set of diagnostic functions. This command is equivalent to selecting
Instrument Diagnostics from the Utilities menu and then enabling Halt on Fail.
DIAg:CONTROL:LOOP
DIAg:CONTROL:HALT (ON|OFF|<NR1>)
ON enables the halt function, causing the execution of diagnostics to halt at the
first diagnostic failure that occurs.
OFF disables the halt function, allowing the instrument to execute the entire set
of diagnostics before halting, even if a diagnostic failure occurs.
<NR1> A 0 enables the halt function; any other number disables the halt
function.
DIAg:CONTROL:HALT ON
This command enables the halt function, causing the execution of diagnostics to
halt at the first diagnostic failure that occurs.
Description
Terminates the execution of diagnostics
Selects and executes any item at any level of the
test hierarchy
Returns a brief pass or fail status of the last test
execution
Returns a more explanatory message about the
results of the last test execution
Terminates the execution of the test