Option VGA Remote Commands
ReportGenerate <pathstring>
Syntax
Group
Arguments
Returns
Examples
3- 504
Generates a measurement report of the specified type (if a measure has been run
and results are available), and saves it in the file specified by pathstring.
You have to specify the extension for the filename.The command does not
append the filename extension automatically. The type of file (.rtf, .csv or .pdf) is
determined by the file extension in the pathstring. If the extension does not
match one of these endings, an error is returned.
Query returns "OK" unless the command is still being processed, in which case
it returns the current pathstring argument.
If you get the error message "Invalid Filename," confirm that you typed the
correct path and that your path is no more than 60 characters.
VARIable:VALue "ReportGenerate", "<pathstring>"
VARIable:VALue? "ReportGenerate"
Reporting
<pathstring> path/filename where file is to be stored. Can either be the full path
and filename with extension or just the filename with extension.
Query returns "OK" unless the command is still being processed, in which case
it returns the current pathstring argument.
VARIable:VALue "ReportGenerate", "C:\colorbars.csv"
VARIable:VALue? "ReportGenerate"
Query may return: "ReportGenerate OK"