Performance Verification
54
After you rename Meas 1 to Meas 4, when you take measurements, the data
displays on the right side of the oscilloscope screen, from Extinction Ratio (dB)
down to Average Optical Power (dBm). (See Figure 19). This measurement
order is repeated left to right in the rows in Tables 19 through 21 in the Test
Record.
4. Select the Acq tab to set the acquisition parameters.
a. In the Stop After box, select Condition.
b. In the Condition field, select Number of Acquisitions from the drop-
down menu.
c. Enter 300 (# of acquisitions) in the field below the Condition field.
d. In the Stop Action box, select None.
5. Close the Setup window.
6. Press the Run/Stop button on the oscilloscope to begin a measurement.
The oscilloscope displays an eye pattern as shown in Figure 19.

Figure 19: Measurements taken from the eye pattern

OI1125 E/O Transmitter Instruction Manual
Mask hits
Measurement data