Table 8: Electromagnetic Compatibility (Emc) - Tektronix OI1125 Instruction Manual

Table of Contents

Advertisement

Specifications

Table 8: Electromagnetic compatibility (EMC)

Specification
Emissions
Immunity, enclosure, electrostatic discharge
(ESD)
Immunity, fast transients, electrical
Immunity, AC power line source voltage
dips and interruptions
Immunity, AC power line transients
22
Description
Emissions shall be within the limits specified by the following requirements.
Enclosure:
EN 55011 Class A limits for radiated emissions
AC Mains:
EN 55011 Class A limits for conducted emissions
Emissions:
EN 61000-3-2 AC power line harmonic emissions
EN 61000-3-2 AC power line fluctuation
AS/NZS 2064, Australian emission standard for Industrial, Scientific, and
Medical Equipment
Use only high quality shielded interface cables having a reliable, continuous outer shield
(braid and foil) that has low impedance connections to shielded connector housings at
both ends.
Up to 8 kV air discharge and 4 kV contact discharge with no change to control settings,
no impairment of normal operation, and no damage that prevents recovery of normal
operation by the user.
No loss of stored data, change to control settings, degradation of performance, or
temporary loss of function will occur when the system is subjected to the transients as
described below:
Port
Peak Voltage (kV)
Signal and Control
0.5
AC Power
1.0
IEC 61000-4-4
Voltage Dips:
30% reduction/10 ms, as per EN 61000-4-11
60% reduction/100 ms, as per EN 61000-4-11
Voltage Interruptions:
u95% reduction/5 seconds as per EN 61000-4-11
100% reduction/1 cycle as per EN 61326
No loss of stored data, change to control settings, degradation of performance, or
temporary loss of function will occur when the system is subjected to the transients as
described below:
Mode
Peak Voltage
Common
2 kV
Differential
1 kV
IEC 61000-4-5
Tr/Th (ns)
Rep. Freq. (kHz)
5/50
5
5/50
5
1
OI1125 E/O Transmitter Instruction Manual

Advertisement

Table of Contents
loading

Table of Contents