Prerequisites; Equipment Required; Table 11: Test Equipment - Tektronix OI1125 Instruction Manual

Table of Contents

Advertisement

Performance Verification

Prerequisites

Equipment Required

32
The tests in this section comprise an extensive, valid confirmation of perfor-
mance and functionality when the following requirements are met:
H The instrument must have been last calibrated and adjusted at an ambient
temperature between +20 _C and +30 _C, must have been operating for a
warm-up period of at least 30 minutes, and must be operating in an
environment with temperature, altitude, humidity, and vibration within the
operating limits described in Environmental characteristics on page 21.
H The test equipment (oscilloscope, pattern generator, tunable laser and optical
meters), must be operating for a warm-up period of at least 30 minutes.
Table 11 lists the equipment required to perform the performance verification
procedure. The types and quantities of connectors may vary depending on the
specific equipment you use.
CAUTION. To prevent damage to your equipment and cables, use the proper type
optical connectors. The optical connection styles shown in the test setups are
correct for the equipment called out in Table 11 below. If you use alternative
equipment, verify the connection style of the alternative equipment, and use the
appropriate cable or adapter.

Table 11: Test equipment

Description
Minimum requirements
Sampling oscilloscope
Bandwidth: 20 GHz with limiting filters
with optical sampling
for OC 48, OC192, and 12.5 Gb/s
Jitter: t1 ps at min samp delay
head
t2.4 Gb/s to u12.5 Gb/s,
Pattern generator
PRBS 2
Wavelength meter
1530 nm to 1565 nm
Optical power meter with
1550 nm, +15 dBm to –40 dBm,
0.01 dBm resolution, ±5% accuracy
optical power head
Tunable Laser
10 dBm output power over 1530 nm to
1565 nm range,
PMF-style output connector
50 W, SMA male each end, ~1 m
Cable, low-loss coax (2)
50 W, SMA male each end, 1 m
Cable, coax
31–1
OI1125 E/O Transmitter Instruction Manual
Example product
Tektronix TDS8000 with
80C01 optical sampling
head
Advantest D3186 with
options 13 and 72
Burleigh WA-1100
Advantest Q8221 with
Q82227 power head
Tunics BT-1560 w/high
power option
Tektronix part number
015-0561-00
Tektronix part number
174-1341-00