Table 7: Environmental characteristics

Specification
Temperature, nonoperating
Temperature, operating
Humidity, nonoperating
Humidity, operating
Random vibration, nonoperating
Random vibration, operating
Packaged product vibration and shock
Altitude, operating and nonoperating
Weight
instrument
only
shipping
Dimensions
The humidity limits are derived from a psychrometric chart using a maximum wet bulb temperature of 29 _C. The highest
1
RH (20%) at the maximum temperature (+60 _C) diverges from the psychrometric chart (which would indicate 6%). The
20% number is a practical limit. Test chambers cannot be set below this number.
OI1125 E/O Transmitter Instruction Manual
Description
(–55 _C to +75 _C)
The environmental exposure is the procedure stated in Tektronix Design Standard
062-2847-00 for Class 6 equipment.
0 _C to +40 _C (32 _F to 104 _F)
This procedure conforms to the temperature portion of the test in MIL-T-28800E for
Type III, Class 6 equipment, using the limits above.
0% to 75% relative humidity from +30 _C to +60 _C.
0% to 80% relative humidity to +31 _C, decreasing at a linear rate to 50% at +40 _C,
minimum.
2.46 g
, from 5 to 500 Hz, 10 minutes each axis.
RMS
Tektronix Standard 062-2858-00 rev. B modified, Random Vibration, Tektronix Class 3
with power spectral density break-points as follows:
2
Constant 0.020 g
/Hz from 5 Hz to 100 Hz, then falls at –3 dB/octave from 100 to
2
200 Hz, then constant 0.010 g
/Hz from 200 to 350 Hz, then falls at –3 dB/octave from
2
350 to 500 Hz to 0.007 g
/Hz @ 500 Hz.
0.31 g
, from 5 to 500 Hz, 10 minutes each axis.
RMS
Tektronix Standard 062-2858-00 rev. B, Random Vibration, Tektronix Class 3
The packaged product qualifies under the Distribution Cycle 1 Assurance Level II for
packaged products 0 to 20 lbs. Test 2 for Warehouse and Vehicle Stacking (compression)
is omitted.
Tektronix standard 062-2858-00, Rev. B.
Operating:
3000 m (9847 ft)
Nonoperating:
12190 m (40,000 ft)
MIL-T-2880E modified for more exposure time and more severe test levels than required
for any Type III equipment regardless of Class.
1.72 kg (3.79 lb.)
5.22 kg (11.5 lb.), including standard accessories
45 mm x 204 mm x 331 mm (1.75 in x 8.0 in x 13 in) (H x W x D)