Laser Safety Label; Figure 10: Laser Warning Label; Table 9: Safety Characteristics - Tektronix OI1125 Instruction Manual

Table of Contents

Advertisement

Table 8: Electromagnetic compatibility (EMC) (cont.)
Specification
Immunity, conducted disturbances induced
by RF fields
1
2/50 us Tr/Th voltage into open circuit, 8/20 us Tr/Th current into short circuit.

Table 9: Safety characteristics

Specification
Power consumption
Source frequency
Safety certifications
EC Declaration of Conformity - Low Voltage

Laser Safety Label

OI1125 E/O Transmitter Instruction Manual
Description
No instrument failures when the instrument power leads are injected with a 150 kHz to
80 MHz, 3 V
signal 80% amplitude modulated at 1 kHz.
RMS
IEC 61000-4-6
Description
<40 W when powered from AC line
47 to 63 Hz
UL3111-1, CAN/CSA-C22.2 No. 1010.1
Compliance was demonstrated to the following specification as listed in the Official
Journal of the European Communities:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2:1995
ment control and laboratory use.
The label shown in Figure 10 is affixed to the top of the instrument cabinet. To
prevent personal injury, it is important that you heed all cautions when working
with laser equipment.
INVISIBLE LASER RADIATION
AVOID DIRECT
EXPOSURE TO BEAM
MAXIMUM 20mW at 1550nm
CLASS IIIb LASER PRODUCT
CAUTION
INVISIBLE LASER RADIATION
DO NOT STARE INTO BEAM
INSTRUMENTS
CLASS 3A LASER PRODUCT
MAXIMUM 20mW at 1550nm
IEC 60825-1:1993+A1:1997

Figure 10: Laser warning label

Safety requirements for electrical equipment for measure-
Specifications