T .O . 33K8-4-551-1
APPENDIX A (Cont.)
Frequency
AC Current
Range : 0 to 10 MHz, VAC;
0 to 100 kHz, AAC
Accuracy: 1 to 40 Hz, ±0 .05% of rdg;
40 Hz to 10 MHz, ±0 .01% of rdg
Range : 0 to 1 AAC
Accuracy:
±(% of rdg + % of mg)
Range
Frequency
Absolute
100 µA
10 to 20 Hz
(0.4005 + 0 .03)
20 to 40 Hz
(0 .1501 + 0 .03)
40 to 45 Hz
(0 .1500 + 0 .03)
45 to 100 Hz
(0 .0601 + 0 .03)
100 Hz to 1 kHz
(0 .0601 + 0 .03)
1 mA to
10 to 20 Hz
(0 .4005 + 0 .03)
100 mA
20to40Hz
(0 .1501+0 .03)
40 to 45 Hz
(0 .1500 + 0 .03)
45 to 100 Hz
(0 .0601 + 0 .03)
100 Hz to 5 kHz
(0 .0601 + 0 .03)
5 to 20 kHz *'
(0 .0601 + 0 .02)
20 to 50 kHz *'
(0 .4000 + 0 .04)
50 to 100 kHz *'
(0 .5500 + 0 .15)
IA
10 to 20 Hz
(0.4005 + 0 .02)
20 to 40 Hz
(0 .1601 + 0 .02)
40 to 45 Hz
(0 .1600 + 0 .02)
45 to 100 Hz
(0 .0800 + 0 .02)
100 Hz to 5 kHz
(0 .1000 + 0 .02)
5 to 20 kHz *'
(0 .3000 + 0 .02)
20 to 50 kHz *'
(1 .0000 + 0.04)
*' Typical accuracies, will not be calibrated.
*2 Calibration of the complete Voltage and/or Frequency ranges at these points are not possible due to the
capability limitations of the Calibration Standards . The TI does not require a limited certification for these
Standards limitations.
* 3 Opt HO1 1000 V rms max (All other units 700 V rms max) .