T .O . 33K8-4-551-1
APPENDIX A (Cont .)
AC Voltage
Range : 0 to 1000 VAC
(Random Sampled Mode) (Cont.)
AC BAND 52 MHz
Accuracy:
±(% of rdg + % of mg) + µV
Range
Frequency
Absolute
1V
20 to 40 Hz
(0 .0803 + 0 .002) + 0
40 Hz to 20 kHz
(0 .0800 + 0 .002) + 0
20 to 50 kHz
(0 .0801 + 0 .002) + 0
50 to 100 kHz
(0 .1070 + 0 .002) + 0
100 to 300 kHz
(0 .3004+0 .01)+0
300 to 500 kHz
(1 .0003 +0 .01)+0
500 kHz to 1 MHz
(1 .0018+0 .01)+0
1 to2MHz
(1 .5033+0 .01)+3
10V
20 to 40 Hz
(0 .0803 + 0 .002) + 0
40 Hz to 20 kHz
(0.0800 + 0 .002) + 0
20 to 50 kHz
(0 .0801 + 0 .002) + 0
50 to 100 kHz
(0 .1138+0 .002)+0
100 to 300 kHz
(0 .3006 + 0 .01) + 0
300 to 500 kHz
(1 .0008 + 0 .01) + 0
500 kHz to 1 MHz
(1 .0018+0 .01)+0
1 to 2 MHz
(1 .5033+0 .01)+3
*2
100 V
20 to 40 Hz
(0 .1202 + 0 .002) + 0
40 Hz to 20 kHz
(0 .1200 + 0 .002) + 0
20 to 50 kHz
(0 .1202 + 0 .002) + 0
50 to 100 kHz
(0 .1204 + 0 .002) + 0
100 to 300 kHz
(0 .4006+0.01)+0
300 to 500 kHz
(1 .5008+0.01)+0
500 kHz to 1 MHz
*2
(1 .5012 + 0 .01) + 0
*2
1000V
* 3
20to40Hz
(0 .3002+0 .01)+0
(5700 V)
40 Hz to 20 kHz
(0 .3000 + 0 .01) + 0
(>700 V)
40 Hz to 20 kHz
(0 .3000 + 0 .01) + 0
20 to 50 kHz
*2
(0 .3003 + 0 .01) + 0
50 to 100 kHz
*2
(0 .3041 +0 .01)+0
AC BAND >2 MHz
Range
Accuracy:
±(% of rdg + % of rng) + µV
Frequency
Absolute
10 mV
20 Hz to 20 kHz
(0 .1006 + 0 .05) + 1 .3
20 to 50 kHz
(0 .1022 + 0 .05) + 2
50 to 100 kHz
(0 .1047 + 0 .05) + 2 .5
100 to 300 kHz
(1 .2027 + 0 .05) + 4
300 to 500 kHz
(1 .2033 + 0 .05) + 8
500 kHz to 1 MHz
(1 .2120+0 .05)+8
1 to 2 MHz
(7 .0007 + 0 .07) + 4 .3
2to4MHz
(7 .0029 + 0 .07) + 4 .3
4to8MHz
(20 .0010 + 0 .08) + 4 .3
See footnotes at end of Appendix A.