T.O . 33K8-4-551-1
4 .4 .6 Set the Calibrator for the first value listed in the Applied column of Table 8 and set the OPR/STBY switch to
OPR.
4 .4 .7 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column of Table 8.
4 .4 .8 The TI display must indicate within the corresponding values listed in the Limits column of Table 8.
4 .4 .9 Set the Calibrator OPR/STBY switch to STBY.
4 .4 .10 Repeat steps 4 .4 .4 through 4 .4 .9 for each remaining range and applied value listed in Table 8.
NOTE
Grounding/Shielding variations may be required to obtain a stable and accurate
TI indication . The TI LFILTER may be used in Table 8 for frequencies up
through 50 kHz . Disengage TI LFILTER for frequencies above 50 kHz . To
engage TI LFILTER press the TI keys in the following sequence:
1
Eller
Table 8.
Range (AC)
Applied (VAC)
Limits (VAC)
10 mV
10m@20Hz
9 .99215 to 10 .00785 m
l0m@1kHz
9 .99532 to 10.00468 m
10m@20kHz
9 .99440 to 10.00560 m
10m@50 kHz
9 .98668 to 10.01332 m
10m@ 100kHz
9 .94630 to 10 .05370 m
l00 mV
100m@20Hz
99 .9835 to 100 .0165 m
100m@1kHz
99 .9885 to 100 .0115 m
100m@20kHz
99 .9820 to 100 .0180 m
100m@50 kHz
99 .9652 to 100 .0348 m
l00 m @ 100 kHz
99 .9139 to 100 .0861 m
1V
1 @20 Hz
0 .999864 to 1 .000136
1 @ 1 kHz *)
0 .999906 to 1 .000094
1 @ 20 kHz
0 .999838 to 1 .000162
1 @50kHz
0 .999676 to 1 .000324
1 @ 100 kHz
0 .999177 to 1 .000823
See footnote at end of Table .
L
$H FT)
s1E)
26