HP 3458A Technical Manual page 18

Digital voltmeter
Hide thumbs Also See for 3458A:

Advertisement

T .O . 33K8-4-551-1
Noun
Minimum Use
Specifications
Calibration
Sub
Equipment
Item
2 .12 NULL DETECTOR
2 .13 DC VOLTAGE
DIVIDER
2 .14 SYNTHESIZED
SIGNAL GENERATOR
2 .15 PARALLEL
COMPENSATION
NETWORK
2 .16 TERMINATION
I
2 .17 NANO VOLT/
MICRO OHM METER
Range : 0 to 1000 VDC
Accuracy : N/A
Range : 0 to .8000000
Terminal Linearity : ±0 .2 ppm *2
Range : 2 Hz to 10 MHz
Accuracy : ±0 .0025% of setting
Range : N/A
Accuracy : N/A
Range : 50 Q
Accuracy : NIA
Range : 0 VDC ; 0 Cl
Accuracy : N/A VDC ; ±2 ppm S2
Fluke
845AB
Fluke
720A
Hewlett-Packard
3325B
ESI
PC 101
As Available
Agilent
34420A
*' The following TAR is calculated using the RSS accuracy of the DC Reference Standard and DC Reference
Divider. For high stability TIs, 1 (TAR 3 .5 :1) and 10 VDC range (TAR 3 :1).
*2
The following TAR is calculated using the RSS accuracy of the DC Reference Standard and DC Voltage
Divider . For high stability TIs, 10 VDC range (TAR 3 :1).
*5 The following TAR is calculated using the RSS accuracy of the Digital Voltmeter and Resistance
Standard. For TI 1 mA and 10 mA DC ranges (TAR 3 .6 :1).
*
4 The following TAR is calculated using the RSS accuracy of the Digital Voltmeter and Resistance
Standard. For Tl 100 µA DC range (TAR 3 :1), 1 mA and 10 mA DC ranges (TAR 2 .7 :1).
* 5 See step 3 .9.
3 PRELIMINARY OPERATIONS:
3 .1 Review and become familiar with the entire procedure before beginning the Calibration Process.
WARNING
Unless otherwise designated, and prior to beginning the Calibration Process,
ensure that all test equipment voltage and/or current outputs are set to zero (0) or
turned off, where applicable . Ensure that all equipment switches are set to the
proper position before making connections or applying power.
3 .2 Use low-thermal, shielded leads with spade-lug connectors whenever possible during the Calibration Process.