HP 3458A Technical Manual page 15

Digital voltmeter
Hide thumbs Also See for 3458A:

Advertisement

T.O . 33K8-4-551-1
2 EQUIPMENT REQUIREMENTS:
Noun
Minimum Use
Specifications
Calibration
Equipment
Sub
Item
2 .1
DC REFERENCE
Output : 10 VDC
Fluke
STANDARD
734A/AF
2 .2
DC REFERENCE
Accuracy : ±1 .5 ppm of output
*
I .
*
2
Fluke
Ratio : 10 :1, ±0 .2 ppm ; *I
DIVIDER
* I
100 :1, ±0 .5 ppm
752A
2 .3
SHORTING BARS
Range : N/A
ESI
(2 EA)
SB 103
Accuracy : N/A
Range : 0 to I VDC
Hewlett-Packard
H-P
2 .4 DIGITAL
VOLTMETER
3458A OPT 002
3458A
Accuracy:
±(4 .277 ppm of rdg + 1 ppm of mg) * 3
for 3458A OPT 002;
±(8 .142 ppm of rdg + 1 ppm of mg) * 4
for 3458A
2 .5
2 .6
CALIBRATOR
AC MEASUREMENT
STANDARD
Range : 0 to 1000 VDC and VAC,
20 Hz to 10 MHz ; 1 VAC at 3, 7
and 10 MHz; 0 to 1 ADC
Accuracy : VDC and ADC, N/A;
1 VAC at 3, 7 and 10 MHz,
±(0 .2% of output + 3 µV)
Range : 0 to 1000 VAC
Accuracy : ±(ppm of rdg + µV)
Fluke
5700A
W/5725A
Fluke
5790A
Range
Frequency
±(ppm of rdg + ]AV)
22 mV
20 to 40 Hz
±(190 + 1 .3) (TAR 2 .5 :1)
40 Hz to 20 kHz
±(110 + 1 .3) (TAR 2 :1)
20 to 50 kHz
±(210 + 2) (TAR 3 .3 :1)
50 to 100 kHz
±(310 + 2 .5)
100 to 300 kHz
±(810 + 4)
See footnotes at end of Equipment Requirements .
13