Warnings and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meaning and should be carefully reviewed: Warning Indicates risks to the safety of the patient or user. Failure to follow warnings may result in injury to the patient or user. Caution Indicates risks to the equipment.
Page 6
10. Do not attempt internal repairs or adjustments not specifically detailed in this operating manual. Ensure that readjustments, modifications, and/ or repairs are carried out by persons authorized by Stryker Endoscopy. 11. Use appropriate caution to prevent contact with fluids if the unit is being used with a power supply in patient environments.
Page 7
15. Do not operate with the glass display screen facing downward. 16. Keep the unit away from equipment that uses strong magnets (i.e., large loudspeakers). 17. Do not touch the patient with signal input or output connectors. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or IEC 60601-1-1 harmonized national standards or the combination should be evaluated for safety.
Symbol Definitions The following symbols appear on the product, its labeling, or the product packaging. Each symbol carries a special definition, as defined below: Dangerous High Voltage For indoor use only Direct Current DC power control switch Protective Earth Ground This Side Up Wireless Transmission Fragile...
Product Description The WiSe™ 26" HDTV Surgical Display is a high-definition, wide screen LCD surgical display with a maximum resolution of WUXGA (1920 × 1200 at 60 Hz). The display supports various video inputs, including digital RGB, analog RGB, serial digital interface (SDI), component video (YPbPr/RGB), S-video, and C-video.
Manufacturer:FSP Group, Model No:PMP150-14) (1) Hospital-grade AC power cord (4) M4 × 10 mm VESA screws (2) Cable-management clamps Optional Accessories (not shown) Stryker Part Number WiSe™ HDTV Transmitter 0240030971 0240030951 0240030952 2 – 5 pin DC convertor adapter...
Device Features Front panel Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. 1. Power LED Shines green if the display is powered on; blinks amber if the display is in standby mode.
Rear panel 1. VESA mounting Provide access points for mounting the monitor. holes (100mm) 2. Power switch (hard) Powers the input DC power ON and OFF. 3. Power connector Connects to the 24V DC power converter. 4. Cable-management Organizes cables. clamps 5.
Setup and Interconnection Stryker Endoscopy considers instructional training, or inservice, an integral part of this device. Your local Stryker Endoscopy sales representative will perform at least one inservice at your convenience to help set up your equipment and instruct you and your staff on its operation and maintenance. To schedule an inservice, contact your local Stryker Endoscopy representative after your equipment has arrived.
Connecting AC Power Connect the power supply to the 24V input on the display. Connect the power cord to the power supply. Connect the AC power, using the supplied hospital-grade power cord. (Optional, not shown) Connect an extension cord between the power supply and the display.
Basic Video Setup WiSe™ HDTV Surgical Display SDC or SIDNE VCR or Printer Camera Route the video output 1 from the camera to the SDC or SIDNE. Route the video output 1 from SDC or SIDNE to input on the DVI input on the display.
Operating the Display Accessing the On-Screen Display Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. 1. Power LED Indicates menu current status. Displays green if display is powered on; blinks amber if display is in Standby mode.
While the OSD menu is deactivated, rotate the Rotary Control to activate the input signal selection menu. The current input signal will be indicated by a dot. Rotate the Rotary Control to select the preferred input signal. Stryker Camera Preset Modes Horizontal Vertical Resolution...
On-Screen Display Menus Item Function Description Range Specialty Color Temperature * Chooses between color temperatures for Standard, Arth, Lap, PACS, or Norm Red balance -128 – 128 Green Green balance -128 – 128 Blue Blue balance -128 – 128 Gamma Gamma value 0.1 –...
Page 19
Item Function Description Range Auto Source Select Adjusts Auto Source Select between on and off Restore Factory Sets to factory default Settings Key Lock Sets to key lock mode Wireless Mac ID Unique machine ID of WiSe™ HDTV Transmitter Status Status message: Receiver: RX MAC: XXXXXXXX...
Using the Optional Fiber Optic Module The display features an optional optical module accessory (0240030962), which allows it to receive a high-definition video signal over fiber optic cables. Installing and Activating the Fiber Optic Module Caution Excessive force or misalignment during insertion might damage the module connector.
Using the Optional WiSe™ HDTV Transmitter The display can be used with the optional WiSe™ HDTV Transmitter (0240030971), which allows it to receive a high-definition video signal over a radio-frequency link. This link is established by means of the token, which is included with the transmitter.
Rear Panel 1. DVI Input Connects to camera signal. 2. DVI Output Connects to primary display. 3. RS232 Port Maintenance port (not for customer use). 4. Equipotential Ground Plug 5. AC Power Inlet Connects to separable power cord that can be used for mains isolation.
Setting Up the Optional WiSe™ HDTV Transmitter Warning When the transmitter is used with other equipment, leakage currents may be additive. Ensure that all systems are installed according to the requirements of IEC 60601-1-1. Always set up the transmitter in a location that allows adequate ventilation.
Route the video output 1 from the camera to the input on the SDC or SIDNE. Route the video output 1 from the SDC or SIDNE to the input on the transmitter. Connect the DVI output from the transmitter to the primary display. Connect the C-video input on the display to the C-video output on a VCR or printer.
Page 25
Power on the transmitter and auxiliary display. Insert the token into the token slot on the transmitter. The token LED shines amber as it writes the data. When the token LED turns green, remove the token from the transmitter. Within 2 minutes, insert the token into the token slot on the auxiliary display.
Cleaning and Maintenance Caution Do not expose the display to moisture or apply liquid cleaners directly to the screen. Spray the cleaning solution into a soft cloth and clean the screen gently. No specific liquid or chemical is necessary for cleaning the display. Use only non-abrasive cloths and cleaning solutions to clean similar equipment used in hospitals.
<0.1 ohms, and power consumption less than or equal to rated power. Use a true RMS digital multimeter and safety analyzer to perform this test. Note Refer calibration and operating difficulties not detailed in this manual to your Stryker Endoscopy sales representative.
Disposal This product contains electrical waste or electronic equipment. It must not be disposed of as unsorted municipal waste and must be collected separately in accordance with applicable national or institutional related policies relating to obsolete electronic equipment. Dispose of any system accessories according to normal institutional practice relating to potentially contaminated items.
Troubleshooting Before returning your display for service, consult the troubleshooting list below: Problem Current Status Remedy No picture Power LED on Using the OSD, adjust the brightness and contrast to maximum, or reset them to their default settings. Power LED off Ensure the power switch at the front and back of the display are set to Check if the AC power cord is...
Technical Specifications Display LCD Display Panel 64.9 cm (a-Si TFT active matrix LCD) Synchronization 2.5 – 5.0 Vpp separated sync Pixel Pitch 0.2865(W) × 0.2865(H) Response Time <25ms Typ View Angle ±89° (L/R) × ±89° (U/D) Display Colors 16 million colors Native Resolution 1920 dots ×...
Page 31
Classification and Approvals Class I Equipment Medical equipment with respect to electric shock, fire, and mechanical hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1. IPX1 Water Ingress Protection Continuous operation Optional WiSe™ HDTV Transmitter Video Input Digital: One Digital Video Interface (DVI) Connector:...
Page 32
Complies with FCC Regulations FCC 15B 2008 (Class B) FCC Identifier: VQSAMN11100R44 Please contact your local Stryker Endoscopy sales representative for information on changes and new products. Products referenced with ™ designation are trademarks of Stryker. Products referenced with ® designation are registered trademarks of Stryker.
Electromagnetic Compatibility Warning When this device is connected with other electrical equipment, leakage currents may be additive. To minimize total leakage current per patient, ensure that all systems are installed according to the requirements of IEC 60601-1-1. Caution Portable and mobile RF communications equipment may affect the normal function of the display.
Page 34
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions The WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter is intended for use in the electromagnetic environment specified below. The customer or the user of the WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic Environment - guidance...
Page 35
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity The WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter is intended for use in the electromagnetic environment specified below. The customer or the user of the WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter should ensure that it is used in such an environment. IEC 60601 Test Compliance Electromagnetic Environment--...
Page 36
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity The WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter is intended for use in the electromagnetic environment specified below. The customer or the user of the WiSe™ 26" HDTV Surgical Display and WiSe™ HDTV Transmitter should ensure that it is used in such an environment. IEC 60601 Test Compliance Electromagnetic...
Page 37
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the WiSe™ 26" HDTV Surgical Display The WiSe™ 26" HDTV Surgical Display is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the WiSe™ 26" HDTV Surgical Display can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the WiSe™...
Stryker Endoscopy. If a valid warranty claim is received within the Warranty Period, Stryker will, in its sole discretion: (1) repair the product at no charge, (2) replace the product at no charge with a product that is at least functionally equivalent to the original product, or (3) refund the purchase price of the product.
Page 39
Stryker’s obligation under this warranty, and such warranty shall survive inspection, test, acceptance, and use. Notwithstanding the above, the following products are warranted for a period of ninety (90) days from the date of purchase: Scopes, Associated Scope Hardware, Fiber Optic Cables, Laparoscopic Instruments, VCRs, Monitors, and Printers;...
Page 40
If a return does not comply with these terms, Stryker Endoscopy reserves the right to destroy the product at the customer’s expense. Any replacement would be at the customer’s expense.
Page 42
Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 USA 1-408-754-2000, 1-800-624-4422 www.stryker.com European Representative: Regulatory Manager, Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex, France 1000401089 B 2008/11...