EXCELL PRECISION CO., LTD.
3-2-1 External Weight Calibration
Zero value can be calibrated separately from the weight calibration value.
Press
key
MC
.
No weight display
Make sure that no weight
on platter and press
key to read "Zero"
.
Calibration poises value
~
Use
key to modify
0
9
calibration poises value,
place enough poises accord-
ing to value entered, and
press
key to read
MC
calibrated internal value and
save into EEPROM.
.
Poises weight value on platter
In weight calibration mode,
Put on or remove weight to
test. After testing, press
key to complete external
weight calibration.
03004035
X X X
If no temperature IC, "notMP" displays in UNIT WEIGIH
kg
X X X
A/D value of temperature
Press
key to quit "zero" calibration
MC
kg
X X X
Calibration poises value entered can not be higher than maximum
weighing capacity.
Poises weight entered will be compared with internal calibration
poises weight. If the actual weight is more than the entered poised
weight in 0.9~1.1 times, "Error" displays for 1 second in WEIGHT.
Re-enter the calibration poises value, or place the correct weight
to operate once again.
If not to calibrate weight, press
value into EEPROM.
kg
X X X
MC
X X X X X X
Internal value
X X X X X X
Internal value
key to exit without saving A/D