EXCELL PRECISION CO., LTD.
3-1-7 Zero Tracking Range Setting
Press
key
MC
Enter parameter 00~03
Press
key to
MC
save the setting
When CFN 01 = 00 ~ 01, the default setting is
When CFN 01 = 00 ~ 04, the default setting is
After weight keeps stable for over 1 second, it could track 1/4 d.
After weight keeps stable for over 1 second, it could track 1/2 d.
After weight keeps stable for over 1 second, it could track 1 d.
After weight keeps stable for over 1 second, it could track 2 d. (d=division)
Only when gross = 0, zero tracking function is activated.
3-1-8 Accumulation Ending Mode Setting
Press
key
MC
Enter parameter 00~02
Press
key to
MC
save the setting
Default Setting:
Press
key. After screen displays accumulation
M+
value for 3 seconds, scale returns to weighing mode.
Press
key. And screen displays accumulation
M+
value, scale does not return to weighing mode until
pressing
CE
Press
key. And screen does not display
M+
accumulation value, but beeper beeps once.
03004035
Previous setting
Previous setting
key
.
.
.
Exit
11
Move cursor leftward
CE
1
Move cursor rightward
T
1
Enter
MC
1
20
ZSME200000058