Appendix B: Specifications; System Battery; X-Ray Source; Focus - GE Optima XR220amx Operator's Manual

X-ray system
Hide thumbs Also See for Optima XR220amx:
Table of Contents

Advertisement

DRAFT

Appendix B: Specifications

Power
x
Capable of 100-240V nominal, 50 /60 Hz operation
x
System battery status display

System Battery

x
Independent component power management and system battery charging algorithms to maximize
usage time
x
System Battery charge can take 90 typical exams on a single charge
x
Maximum charge time of 4 hours to go from 10-100%
X-ray Tube

X-ray Source

x
Heat Storage Capacity
– Anode: TBD Joules (TBD heat units)
– Tube Unit: TBD Joules (TBD heat units)
x
Maximum Heat Dissipation Rate
– Maximum anode heat units: TBD
– Maximum Continuous Heat Dissipation: Without Air-circulator: 170W (238 HU/s)
x
Focal Spot Sizes 1.3mm/0.6mm
x
Target Angle 12.5°
x
Anode Rotation Speed (minimal): 3200/min
x
Maximum X-ray Tube Current
– Large Focus: 500 mA
– Small Focus: 200 mA
x
Nominal Voltage (Radiographic): 40 to 150 kV
x
Maximum Voltage:
– Anode to Cathode TBD kVp rectified
– Anode or Cathode to ground TBD kVp
Optima XR220amx X-Ray System