Page 2
1) SAFETY This manual contains information and warnings that must be followed for operating the meter safely and maintaining the meter in a safe operating condition. If the meter is used in a manner not specified by the manufacturer, the protection provided by the meter may be impaired.
UL Listed Probe Assembly to the same meter ratings or better. CAUTION Disconnect the test leads from the test points before changing functions. International Electrical Symbols Marking of Electrical and Electronic Equipment (EEE). Do not dispose of this product as unsorted municipal waste. Contact a qualified recycler Caution! Refer to the explanation in this Manual Caution! Possibility of electric shock Earth (Ground)
3) PRODUCT DESCRIPTION Note: Top of the line model is used as representative for illustration purposes. Please refer to your particular model for function availability. 1) 3-5/6 digits 6000 counts dual displays 2) Push-buttons for special functions & features 3) Selector to turn the Power On or Off and Select a function 4) Input Jack for Insulation...
4) OPERATION CAUTION Before and after hazardous voltage measurements, test the voltage function on a known source such as line voltage to determine proper meter functioning. & VFD ACV functions Press the SELECT button momentarily to toggle the subject functions. Last selection will be saved as power up default for repeat measurement convenience.
Page 6
Resistance, Continuity functions Press the SELECT button momentarily to toggle the functions. Last selection will be saved as power up default for repeat measurement convenience. Continuity function is convenient for checking wiring connections and operation of switches. A continuous beep tone indicates a complete wire.
Page 7
The compensation value stays until the next meter power reset, and can be as high as 5. If you need a compensation value that is higher than that, Relative Zero mode is recommended. Diode Test, Capacitance (Model 887 only) functions Press the SELECT button momentarily to toggle the functions.
Page 8
DCmV, ACmV functions Press the SELECT button momentarily to toggle the subject functions. Last selection will be saved as power up default for repeat measurement convenience.
Page 9
Temperature functions (Model 887 only) Press SELECT button momentarily to toggle C and F readings. Last selection will be saved as power up default for repeat measurement convenience. Note: Be sure to insert the banana plug type-K temperature bead probe Bkp60 with correct polarities.
Page 10
A, mA, µA Current functions Press the SELECT button momentarily to toggle the DC or AC of the subject functions. Last selection will be saved as power up default for repeat measurement convenience. For Model 885, A and mA functions are at different Rotary Switch positions. Insulation Resistance function WARNING The two LCD icons...
Page 11
power from the circuits before carrying out the Setup as illustrated below. Press the RANGE (Test Voltage) button momentarily to select an intended test voltage of 50V, 100V, 250V, 500V or 1000V in sequence. Last selection will be saved as power up default for repeat measurement convenience.
Page 12
Lock-Test mode is recommended for continuous measurements. To apply, press the LOCK button momentarily to display the annunciator before pressing the TEST button momentarily. The LCD will show both & to indicate continuous measurement is active. Press again either button momentarily to release the Lock-Test mode.
Page 14
Smooth mode (Insulation Resistance function only) Smooth mode displays the running average of the last eight measured readings having changes within 300 counts in sequence. On the contrary, it displays directly, without smoothing, the measured reading that is beyond 300 counts in changes comparing to its former one.
other functions, especially voltage function, is selected. Set Beeper Off Press the RANGE button while turning the meter on to temporarily disable the Beeper feature. Turn the rotary switch OFF and then back on to resume. Auto-Power-off (APO) The Auto-Power-off (APO) mode turns the meter off automatically to extend battery life after approximately 20 minutes of no rotary switch or push button operations.
Page 16
Battery and Fuse replacement Battery use: Four 1.5V AA alkaline batteries (IEC LR6) Fuses use: Fuse (F2) for “ µA mA” input: 0.4A/1000Vac & Vdc, IR 30kA, F fuse; or better. Dimension: 6 x 32 mm Fuse (F3) for “A” input: 11A/1000Vac &...
Page 17
GENERAL SPECIFICATION Display: 3-5/6 digits 6,000 counts Polarity: Automatic Update Rate: 5 per second nominal 61 Segments Bar graph: 40 per second max Operating Temperature: -10C to 50C Relative Humidity: Maximum relative humidity 90% for temperature up to 28C decreasing linearly to 50% relative humidity at 50C Pollution Degree: 2 IP Rating: IP40 Storage Temperature: -20C to 60C, <...
Page 18
100V output voltage: 45mA 250V output voltage: 85mA 500V output voltage: 170mA 1000V output voltage: 440mA Tester can perform at least 950 insulation tests with new alkaline batteries at room temperature. These are standard tests of 1000 V into 1 M ...
Page 19
AC Voltage RANGE Accuracy Model 887 Model 885 50Hz ~ 60Hz 60.00mV , 600.0mV , 6.000V, 60.00V, 600.0V, 1000V 0.7% + 4d 0.7% + 4d 40Hz ~ 1kHz 60.00mV , 600.0mV , 6.000V, 60.00V, 600.0V 1.3% + 4d 1.3% + 4d 1000V 2% + 4d 2% + 4d...
Page 20
DC Voltage RANGE Accuracy Model 887 Model 885 60.00mV 0.2% + 3d 0.3% + 3d 600.0mV, 6.000V, 60.00V 0.1% + 2d 0.2% + 2d 60.00mV, 600.0V, 1000V 0.2% + 3d 0.3% + 3d Input impedance: 10M, 110pF nominal Ohms RANGE Accuracy Model 887 Model 885...
Page 21
Capacitance (Model 887 only) RANGE Accuracy 1.5% + 5d 2.000F , 20.00F, 200.0F, 2000F 20.00mF 5% + 5d Accuracies with film capacitor or better Specified from 0.200F Temperature (Model 887 only) 1) 2) RANGE Accuracy -40.0C ~ 0.0C 1% + 2C 0.0C ~ 50.0C 2.2C 50.0C ~ 537.0C...
Page 22
AC current RANGE Accuracy Burden Voltage Model 887 Model 885 50Hz ~ 60Hz 600.0A, 6000A 0.2mV/A 1% + 3d 60.00mA, 600.0mA 3mV/mA 6.000A, 10.00A 30mV/A 40Hz ~ 3kHz 600.0A, 6000A 0.2mV/A 2% + 3d 60.00mA, 600.0mA 3mV/mA 6.000A, 10.00A 30mV/A 3kHz ~ 5kHz 600.0A, 6000A 0.2mV/A...
Page 23
Record mode This mode records standard measurement Max and Min readings on most functions, Manual or Auto-ranging where available. Nominal response and accuracy: Same as standard measurements Insulation Resistance Test Voltage Range Test Current Accuracy 3.000M, 30.00M, 55.0M 1mA @50k 100V 3.000M, 30.00M, 110.0M...
BRYMEN TECHNOLOGY CORPORATION. BRYMEN assumes no risk for damage in transit. BRYMEN will, at its option, repair or replace the defective product free of charge. However, if BRYMEN determines that the failure was caused by misused, altered, neglected, or damaged by accident or abnormal conditions of operation or handling, you will be billed for the repair.