Allen-Bradley 1441-PEN25 Release Notes page 9

Enpac 2500 data collector version 3.10
Table of Contents

Advertisement

Table 3 - Corrected Anomalies with Firmware Revisions 1.03...3.10
Operating
Description
System Version
3.10
CORRECTED: Offroute does not take displacement from Accel in Time format.
CORRECTED: Using SD Card, formatted as Fat32, you need to take 2 sets of data for the same route.
CORRECTED: Further to the Orbit issue, there is another problem concerning Gap Voltage measurements from proximity probes.
The data collector always unloads a value of either -4.0882V when collected with an Orbit or -4.0868V when collected as a separate measurement. This is
regardless of the fact that the actual gap voltage is, for example, -9.987 or -10.023.
3.0
CORRECTED: The data collector skipped measurement points with STD Envelope filters.
CORRECTED: The data collector inverted timewave forms. The cause was unique to specific measurement ordering.
CORRECTED: The data collector would not take GSE readings with user defined units.
CORRECTED: When reviewing data, the spectra plots would not be displayed.
CORRECTED: Frequency items would not load when loading routes through the RS232 cable.
CORRECTED: The data collector would not store data from temperature gun.
CORRECTED: The data collector would not turn on ICP power for measurements using velocity sensors. This would cause errors in resulting data.
CORRECTED: When applying notes to spectra, the data collector would "zero" out all amplitudes on the spectrum.
CORRECTED: If you have band alarms that are overlapping then if one is triggered but the overlapping is not, you will not see the whole alarm band flashing as
the untriggered alarm is painted over the triggered one.
CORRECTED: gSE/Time measurements may not be taken correctly; the instrument has displayed all zeros when you follow this sequence.
• Restart the instrument.
• Take a gSE Time Only point.
CORRECTED: When using a Velocity probe, both the data collector, revision 2.02, disables ICP power. When using OffRoute and coupling set to ICP, the unit
does not actually turn on ICP power.
CORRECTED: If connected to a Speed Ref POINT, the Spectrum POINT is downloaded with the wrong speed value. This speed value is based on the previously
uploaded Speed. If this POINT is attached to a Speed Ref POINT, it should update to the newly measured speed and then upload to the Emonitor software with
the new value.
CORRECTED: When you setup an OffRoute gSE POINT with Display Format = Time and 2 kHz filter, with 500 Hz Fmax, the gSE time waveforms are inverted. If
you apply a Burst signal, with a carrier of 1 kHz and Burst of 10 cycles and 50 ms (20 Hz), you may see that the spikes in the time waveform are going negative,
rather than positive.
CORRECTED: The first time you delete a measurement (from Review) you get a TourDB error and it isn't deleted (although the tick mark vanishes), but if you
delete it again it disappears.
CORRECTED: When taking an Orbit measurement, the default view shows the correct rpm. If you change to the Time View using the View Menu, then the
incorrect rpm is shown. The Orbit view is correct and is showing what is actually measured and stored.
CORRECTED: If you take a measurement with Frequency Items where the Speed is taken from a previous POINT, for example rpm, then the Frequency Items are
incorrectly drawn in the FFT. They appear to get correctly drawn after the POINT has been stored, for example, they are correct in Review.
CORRECTED: The Machine Speed 'Auto' setting should default to behave like 'Slow' when the associated rpm has less than 600 rpm. 'Auto' may only behave
like 'Slow' when the speed drops below 200 rpm, but this is too low (3.33 Hz).
Rockwell Automation Publication GMSI10-RN004B-EN-E - August 2012
Enpac 2500 Data Collector Version 3.10 9
ICM00000921