6.2.11 Electrical Calibration Page

CALIBRATION
mV Zero
– – – –
or
Enter
mV FSD
– – –
or
Enter
Temp. Zero
– – –
–°C
or
Enter
Temp. FSD
– – –
–°C
or
Enter
Page header.
Advance to next parameter.
Millivolt Input Zero
Set the millivolt source to –400 mV.
Set the display to '–400' mV.
Store. New value is accepted only when input is stable.
Advance to next parameter.
Millivolt Input Full Scale
Set the millivolt source to +400 mV.
Set the display to '+400'mV.
Store. New value is accepted only when input is stable.
Advance to next parameter.
Temperature Zero
Set the resistance box to 96.09 Ω (resistive equivalent of –10 °C).
Set the display to '–10°C'.
Store. New value is accepted only when input is stable.
Advance to next parameter.
Temperature Full Scale
Set the resistance box to 142.29 Ω (resistive equivalent of 110 °C).
Set the display to '110°C'.
Store. New value is accepted only when input is stable.
Advance to next parameter.