6.2.4 Set Up Input Page
SET UP INPUT
Control Temp
– –
or
Enter
Display as
Enter
Ion Units
mg/kg
Enter
Display Zero
– –
Enter
Display FSD
– –
Enter
or
CURRENT OUTPUT
Page header.
Advance to next parameter.
Control Temperature
–
°C
•
Set the required block control temperature within the range 5 to 45 °C in 0.1 °C
increments.
Store.
Advance to next parameter.
Display Type (Ammonium, Ammonia or Nitrogen)
NH
?
4
Select display as ammonium (NH
N?
or
NH
?
3
Store.
Advance to next parameter.
Ion Units
Select the required display units for ammonia concentration.
mg/l
or
ppm
Store.
Advance to next parameter.
Display Zero
–
•
Set the required value for display within the range 0.05 to 20.00.
Store.
Advance to next parameter.
Display Full Scale
–
•
The full scale value is automatically set at two decades above Display Zero.
Example– If the setting is 1.0 the full scale is automatically set to 100.0.
Store.
Return to top of Set Up Input Page.
or
Advance to Current Output Page.