HP 54501A Operation Reference page 65

Front panel operation reference digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

Oscilloscope Setup
Set up the HP 8116A Pulse/FUnction Generator:
=
Mode
NORM
=
FRO
MHz
1.00
=
DTY
50%
=
AMP
1.00 V
=
OFS
-2.00 V
Pulse ;;;: : squarewave
Connect the signal to a BNC tee on channel 1 using a 1 meter coaxial
cable. Connect another 1 meter cable from the other side of the BNC tee
to channel 4.
The extra cable length between channels 1 and 4 provides a
between the signals displayed on the oscilloscope. The propagation of a 1
metre coaxial cable is approximately 6 to 7 ns. This time delay is used to
demonstrate the
HP
5450lA triggering capability.
Press AUTOSCALE.
Set up the HP 5450lA as follows:
=
Timebase
10.00 ns/div
=
delay
0.00 s
=
reference
cntr
=
window
off
Channel l
Vertical sensitivity
=
offset
-200.00 mV
dc coupling
Channel 4
Vertica1 sensitivity
=
offset
--200.00 m V
dc coupling
=
500 mV/div
=
500 mV/div
time
delay
Trigger Menu
6-11