Fluke 233 Calibration Information Manual page 5

True-rms remote display
Hide thumbs Also See for 233:
Table of Contents

Advertisement

Temperature
Range
-40 °C to +400 °C
-40 °F to +752 °F
[1]
Temperature uncertainty (accuracy) does not include error of the thermocouple probe.
AC Current
Function
Range
[1,2,3]
A ac
6.000 A
10.00 A
[1]
All ranges are specified from 5 % of range to 100 % of range.
Crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale.
[2]
[3]
AC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.
DC Current
Function
Range
[1]
A dc
6.000 A
10.00 A
[1]
DC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum.
Capacitance
Range
1000 nF
10.00 μF
100.0 μF
9999 μF
>1000 μF: 5 % + 20
[1]
Diode
Range
2.000 V
Frequency
AC coupled, 5 Hz to 50 kHz, for V ac; dc coupled, 45 Hz to 5 kHz for A ac switch position.
Range
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
Resolution
0.1 °C
0.1 °F
Resolution
0.001 A
0.01 A
Resolution
0.001 A
0.01 A
Resolution
1 nF
0.01 μF
0.1 μF
1 μF
Resolution
0.001 V
Resolution
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
True-rms Remote Display Digital Multimeter
[1]
Accuracy
±(1.0 % + 10)
±(1.0 % + 18)
Accuracy
(45 – 500 Hz)
±(1.5 % + 3)
Accuracy
±(1.0 % + 3)
Accuracy
[1]
±(1.9 % + 2)
Accuracy
±(0.9 % + 2)
Accuracy
±(0.1 % + 2)
Detailed Specifications
5

Hide quick links:

Advertisement

Table of Contents
loading