Fluke 233 Calibration Information Manual page 18

True-rms remote display
Hide thumbs Also See for 233:
Table of Contents

Advertisement

233
Calibration Information
Rotary
Switch
Position
m
C-01
C-02
C-03
C-04
C-05
C-06
C-07
e
C-08
Ohms
C-09
C-10
C-11
C-12
C-13
C-14
K
C-15
C-16
C-17
C-18
C-19
C-20
Set calibrator to standby, reconfigure leads, and program for amps output.
?
C-22
A
C-23
[1]
Wait an additional 5 seconds after calibrator has settled before pressing g.
18
Table 4. Calibration Adjustment Steps
Calibration Steps
Input
Calibrator
Terminals
Source Value
VΩ/+ and COM
0 V, 0 Hz
VΩ/+ and COM
300 mV, 0 Hz
VΩ/+ and COM
100 mV, 0 Hz
VΩ/+ and COM
-300 mV, 0 Hz
VΩ/+ and COM
60 mV, 0 Hz
VΩ/+ and COM
600 mV, 0 Hz
VΩ/+ and COM
600 mV, 60 Hz
600 e, 2-wire comp
VΩ/+ and COM
VΩ/+ and COM
6 ke
VΩ/+ and COM
60 ke
VΩ/+ and COM
600 ke
[1]
VΩ/+ and COM
6 Me
[1]
VΩ/+ and COM
Short
VΩ/+ and COM
40 Me
VΩ/+ and COM
6 V, 60 Hz
VΩ/+ and COM
60 V, 60 Hz
VΩ/+ and COM
600 V, 60 Hz
VΩ/+ and COM
6 V, 0 Hz
VΩ/+ and COM
60 V, 0 Hz
VΩ/+ and COM
600 V, 0 Hz
A and COM
6 A, 60 Hz
A and COM
6 A, 0 Hz
[1]
[1]