Some adjustment steps require additional wait time after the calibrator settles, as noted in
Table 4.
7. After each input value is applied, press g to accept the value and proceed to the next step (C-02 and
so forth).
After pressing g, wait until the step number advances before changing the calibrator
source or turning the Meter's rotary knob. Some adjustment steps can take up to several
seconds to execute before moving to the next step.
If the knob is not in the correct position for a given step, the meter will flash the unit
annunciators until the knob is put in a valid position. The keys that show the reading and
required input values are not allowed until the knob is correct.
Likewise, if the rotary switch is not in the correct position or the measured value is not within
the anticipated range of the input value, the Meter will emit a double beep and will not
continue to the next step when g is pressed.
8. After the final step, the display shows "End" to indicate that the calibration adjustment is complete.
Press g to return to meter mode.
Set the calibrator to Standby prior to changing the function switch position and after
completing adjustment of each function.
If the calibration adjustment procedure is not properly completed, the Meter will not operate
correctly.
True-rms Remote Display Digital Multimeter
Note
Notes
Notes
Calibration Adjustment
17