3.2

Mechanical Specifications

Dimensions: 5.47 x 2.00 x 1.18" (139 x 51 x 30mm)
Weight: 6.5 oz (150g)
Maximum Cable Diameter: 0.78" (20mm)
Jaw Opening: 0.83" (21mm) max
Maximum Conductor Size: 0.78" (20mm)
Envelope Protection: IP 40 (IEC 529)
Drop Test: 1m (IEC 68-2-32)
Mechanical Shock: 100g (IEC 68-2-27)
Vibration: 10/55/10Hz @ 0.15mm (IEC 68-2-6)
Output: 6 ft (2m) insulated lead with insulated BNC Connector
3.3

Environmental Specifications

Operating Temperature/RH: 14° to 131°F (-10° to 55°C); RH <85%
Storage Temperature: -40° to 158°F (-40 to 70°C)
Influence of Temperature: ≤ 0.15% / 10K
Influence of Humidity: 10 to 90%; 0.2%
Altitude: ≤2000m
3.4

Safety Specifications

Indoor Use
Electrical:
Double insulation or reinforced insulation between the primary or second-
ary and the outer case of the handle per EN 61010-2-032.
Common Mode Voltage: 600V, Category III, Pollution Degree: 2
Dielectric Strength: 5550V, 50/60Hz between primary, secondary, and
the outer case of the handle
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50082-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
*Specifications are subject to change without notice