4.2 Exchange of External Bellows

It is not allowed to exchange the internal bellows!
CAUTION
In case that it should be damaged, please send probe for
repair.
Damage of the measuring system
The measuring system is not allowed to be twisted!
TC76-LF (low force):
The bellows is allowed to be exchanged by Blum-Novotest
only!
1. Clean probe and remove stylus if necessary (chap.3.4), while using the
interlocking disk (A) resp. mounting tool (B).
2. Remove front ring (16) with mounting tool (B).
3. Remove external bellows carefully (13), without twisting the measuring
mechanism (10).
4. Grease new bellows (13) at the contact surface to the groove slightly
(grease: ISOFLEX TOPAS NB 52 or STABURAGS NBU30).
5. Install new bellows (13), without twisting the measuring mechanism.
The bellows must click to the groove on the measuring system.
Please take care of plane supporting surface of the bellows.
6. Install front ring (mounting tool (B)).
7. Mount stylus while using the interlocking disk (A) resp. mounting tool (B).
8. Carry out calibration cycle.