3.4 Mounting of the Stylus
CAUTION
Damage of the measuring mechanism
•
•
•
•
1,5 mm
M
= 1 Nm
d
TC60
Fig. 3.4
(A)
interlocking disk (scope of delivery)
(B)
mounting tool (scope of delivery)
© by Blum-Novotest 2010
TC60: When mounting the stylus into the probe, it is
absolutely necessary to lock the probe with the mounting
tool B to avoid torsion of the measuring mechanism.
TC63-30 / TC76:
When mounting the stylus into the probe, it is absolutely
necessary to use the interlocking disk (A) (backlash-free).
Pls. take care of the orientation of the interlocking disk (A)
(top side = top)
Calibrate probe after mounting of stylus!
Damaged styli are not reusable!
TC63-30
Mounting and Commissioning
13