Page 1
Meter correctly. ●Keep this INSTRUCTION MANUAL carefully and refer to this, when necessary. ●In the event of any doubt arising, the original INSTRUCTION MANUAL in Japanese is to be final authority. SANKO ELECTRONIC LABORATORY CO., LTD. (株式会社 サンコウ電子研究所)
CONTENTS ITEM PAGE 1.PRINCIPLE 2.APPLICATIONS 3.SPECIFICATIONS 4.PARTS IDENTIFICATION 5.PREPARATION 5-1. Preparation of the zero plate 5-2. Connecting (Disconnecting) the probe 6.OPERATING INSTRUCTIONS 6-1. Power ON 6-2. Power OFF 6-3. Zero adjustment 6-4. Standard calibration(CAL) 6-5. Resetting the working curve 6-6. Measurement 7.ADDITIONAL FUNCTIONS 7-1.
Before using the Meter, read this INSTRUCTION MANUAL thoroughly and use the Meter correctly. Keep this INSTRUCTION MANUAL carefully and refer to this, when necessary. 1.PRINCIPLE The thickness of non-conductive films treated on non-magnetic metallic substrate can be measured non- destructively, simply, speedily, and accurately utilizing the electric correlation between the eddy current induced on metal surface by a high frequency electric...
3.SPECIFICATIONS ・Name and type Eddy current type digital coating thickness meter EDY-5000 ・Measuring method Eddy current type ・Measuring range 0 to 5.00 mm ±0.01mm on uniform surface or ±2 % of reading ・Accuracy Digital display on large size Liquid Crystal Display (LCD) ・Indication...
4.PARTS IDENTIFICATION ⑤ ⑥ ⑦ ① ② ③ ④ ⑧ ① POWER key Key for power ON/OFF ② ZERO key Key for zero adjustment ③ LOCK key Key for locking all keys except ON/OFF key Key for standard calibration with the standard thickness plate ④...
5. PREPARATION 5-1. Preparation of the zero plate Prepare the substrate which is the same kind, thickness and shape with an object to be measured. The attached zero plate for checking the Meter, Aluminum 1050, is used for the operation check of the Meter. Prepare the substrate of an actual object to be measured.
6. OPERATING INSTRUCTIONS 6-1. Power ON Press the 「ON/OFF」key ① after connecting the probe. [Indication on the LCD ⑦ changes as follows.] ↓ − − − μm Non-Fe Inside of the Meter is being initialized. Keep the probe turn towards the air during this period. ↓...
6-3. Zero adjustment Press the probe against the substrate for adjustment (zero plate). [Indication on the LCD ⑦ changes as follows.] ← The buzzer emits a beeping ↓ sound. XX μm Measured value [XX] is Zero plate indicated on the LCD ⑦. Non-Fe ↓...
6-4. Standard calibration (CAL) Place the standard thickness plate on the zero plate, and press the probe against it. Select a little thicker standard thickness plate than a film thickness to be measured. [Indications on the LCD ⑦ change as follows.] ←...
6-5. Resetting the working curve When the batteries were replaced, indication was locked or measurement, zero adjustment, standard calibration (CAL) became impossible for operation, reset the Meter by following method. The Meter is in power ON condition. 3. 20 mm Non-Fe ↓...
6-6. Measurement Be sure to pass the hand strap ⑧ around the wrist to prevent the Meter from dropping. When the above mentioned zero adjustment and standard calibration are completed, the Meter is ready for the measurement. Hold the mid-part of the probe as shown in left figure and press it against the surface of an object to be measured vertically, quickly and calmly.
8.REPLACING THE BATTERY When the battery closes to the limit of use due to consumption,「LOBAT」is indicated on the lower left of the LCD ⑦. XXX μm As it causes malfunction under this condition, replace the batteries earlier. LOBAT Non-Fe Replace the batteries after surely power OFF. When the batteries are replaced under power ON, (Indication of battery's voltage drop) it causes out of order of the Meter.
10. CARES TO RAISE AN ACCURACY OF THE MEASUREMENT ① Zero plate For use in the zero adjustment and standard calibration(CAL), prepare the zero plate which is same kind, thickness and shape as the substrate of an object to be measured. It will result in an inaccurate measurement to use the zero plate which is different specifications from an object to be measured.
Page 14
Products sold: Sales of Coating thickness meter Pinhole detector Condensator Viscosity cup Moisture meter Needle detector Iron piece detector Manufacturer: Sanko Electronic Laboratory Co., Ltd. ., 2 6 4, Tokyo branch Shibata Bldg Uchikanda Chiyoda-ku 101 0047, Tokyo Japan 81 3 3254 5031 81 3 3254 5038 ., 2 3,...