Calibration of GroundFlex
Calibration is necessary for a new sensor, one that is no longer identified when a
sensor has been replaced, or more than 2 lengths of extension cable is used. The
calibration coefficients are then stored in the Model 6474.
NOTE: Do not move or touch the GroundFlex
during the calibration. This could cause errors.
Start by identifying each sensor using the identification rings that are supplied with
the Model 6474: (1) Brown
Connect the calibration loop between terminals H and E of the Model 6472.
Connect the Model 6472 and 6474 together using the connection cable.
Connect the GroundFlex
6474, the one with the Red ring to Input 2, the one with the Orange ring to Input
3, and the one with the Yellow ring to Input 4.
Coil the GroundFlex
it using its screw connector. Place the connector as far as possible from the
calibration device.