ZERO CALIBRATION
Zero calibration is performed to determine the output of the measuring probe when no light reaches the sensor and to set this as the zero
point to which all other measurements are referenced. Zero calibration must be performed after the POWER switch has been set ON
brfore taking any measurements.
To perform zero calibration :
* Before performing zero calibration, check that the measuring probe has been connected to probe
connector P1.
1. Check that the POWER switch is set to ON.
2. Set the measuring/viewing selector to the
(viewing) position. (An image can be seen in the
viewfinder, but no light will reach the sensor.)
3. Press 0-CAL.
* If zero calibration is being performed immediately
after the POWER switch has been set to ON, press
0-CAL after "PUSH 0-CAL KEY" appears in the
liquid crystal display.
4. Set the measuring/viewing selector to the
Measurements will be started immediately.
1. POWER switch
3. 0-CAL
CA110 Application (Continued)
position.
Display after POWER
DARKEN PROBE
switch is set to ON
Approximately 5 seconds
DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
CH00
P1
E1 [
]
"E1" will appear in the liquid crystal
display the first time the CA-110 is
used after shipment because no
standard color has been set.
Hudson2 200P3
Go to cover page
Press 0-CAL,
Performing zero
calibration
Zero calibration
completed
Measuring/viewing selector