Place the BW100 into the PROGRAM mode at P291.
'B'
'B'
Remove the test weight, insure that the BW100 SW2 is in 'close' position and that P1 is
left as set.
Note:
Performing a balance procedure requires a subsequent zero and span
recalibration.
Page 22
' A '
test weight
' A '
test
weight
Milltronics BW100 - INSTRUCTION MANUAL
test weight on
load cell B side
select P291 ensure SW2
is closed
enter and auto advance to P292,
open SW2
enter and auto advance to P293,
leave SW2 open
test weight on
load cell A side
enter and auto advance to P294,
close SW2
enter and auto advance to P295
SW2
closed
SW2
open
SW2
open
SW2
closed
P1
adjust until
## = 0 +/- 5
7ML19985DJ01