Page 1
LUDLUM MODEL 12-4 SURVEY METER WITH HE-3 NEUTRON DETECTOR May 2016 Serial Number 227878 and Succeeding Serial Numbers...
Page 2
May 2016 Serial Number 227878 and Succeeding Serial Numbers This Manual also applies to Serial Numbers of Model 12-4 Instruments where the BF-3 detector has been modified to He-3. He-3 detectors, and those modified to be, are labeled with an “H.”...
RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Cleaning and Maintenance Precautions Calibration and Maintenance Calibration Establishing an Operating Point Equipment Required for Calibration Initial Calibration Procedures Detector Plateau Gamma Rejection Instrument Calibration Electronic Calibration Maintenance Recalibration Batteries Troubleshooting Troubleshooting Electronics that utilize a Proportional Detector Ludlum Measurements, Inc. May 2016...
Page 5
HV Power Supply Board (Drawing 464 × 243) Switching Convertor Feedback Voltage Control Filtering Recycling Parts List Model 12-4 Survey Meter 10-1 Main Board, Drawing 464 × 275 10-1 HV Power Supply Board, Drawing 464 × 243 10-4 Wiring Diagram, Drawing 464 × 309...
Section 1 Section Introduction he Model 12-4 is a Survey Meter coupled to a Model 42-31H Neutron Detector; providing the required electronic circuitry and detector for measuring and monitoring of neutron radiation. The instrument provides four linear ranges used in combination with an exposure rate or (counts per minute) meter dial.
The Model 12-4 serial number is located on the front panel below the battery compartment. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification.
A mild electric shock may occur if you make contact with the center pin of the input connector. Switch the Model 12-4 range selector switch to the position before connecting or disconnecting the cable or detector.
Model 12-4 Technical Manual Section 2 Instrument Test After checking the batteries, turn the instrument range switch to the ×1000 position. Place the switch in the position. The instrument AUD ON speaker should emit ˝clicks˝ relative to the rate of counts detected. The switch will silence the audible clicks if in the position.
Model 12-4 Technical Manual Section 3 Section Specifications Operating Voltage : approximately 1200 Vdc : 2 Atm He tube – LND 25185 or equivalent – proportional Detector detector, surrounded by a 22.9 cm (9 in.) diameter polyethylene sphere : -2 mV...
Page 11
Model 12-4 Technical Manual Section 3 : reading within 10% of true value with detector connected Linearity : adjustable from 1 to 50 mV Discriminator : built-in unimorph speaker with switch (greater than Audio 60 dB at 61 cm {2 ft})
Model 12-4 Technical Manual Section 4 Section Identification of Controls and Functions : 6.4 cm (2.5 in.) arc, 1 mA analog type with pivot-and-jewel Meter suspension. Typical meter dial is 0-10 mrem/hr, 0-2.5 kV and BAT TEST Connector : Used to connect the detector to the instrument. Typically series ˝C,˝...
Page 13
Model 12-4 Technical Manual Section 4 : When depressed, this button provides a rapid means RES Pushbutton of driving the meter needle to zero. : provides meter response. Selecting the fast, ˝ ˝ position Toggle Switch of the toggle switch provides 90% of full-scale meter deflection in four seconds.
Caution! The operator or responsible party is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc. Caution! Verify instrument voltage input rating before connecting to a power converter.
European Union. Cleaning and Maintenance Precautions The Model 12-4 instrument and detector may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning or performing maintenance on the instrument: 1.
(Discriminator) controls the amplifier gain. Equipment Required for Calibration All instruments used in calibrating the Model 12-4 must be calibrated by standards traceable to the National Institute of Standards and Technology and must have a current calibration label attached. Ludlum Measurements, Inc.
Model 12-4 Technical Manual Section 6 A Ludlum Model 500 Pulser or equivalent is required. If the pulser does not have an HV readout, use a high-impedance voltmeter with at least 1000 Megohm meter input resistance to adjust the detector voltage.
(R49) on the main board until the pointer is over the 1.5 kV mark. Place the Model 12-4 back in the can, with the unimorph speaker on the left side. Depress the Model 12-4 button and adjust the Model 12-4 HV CAL potentiometer (front panel) until the Model 12-4 meter indicates 0.5 kV.
Adjust the Model 500 pulse rate to the calculated value. (Count rate should be 9600 cpm ±10%.) Set the Model 12-4 range selector switch to ×10 and adjust the ×10 calibration potentiometer for a meter reading of ˝8˝.
Maintenance Instrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 12-4 instrument may be cleaned with a damp cloth (using only water as the wetting agent). Do not immerse instrument in any liquid. Observe the following precautions when cleaning: 1.
Check the appropriate regulations to determine required recalibration intervals. Ludlum Measurements offers a full-service repair and calibration department. We not only repair and calibrate our own instruments, but most other manufacturers’ instruments. Calibration procedures are available upon request for customers who choose to calibrate their own instruments.
Note that the first troubleshooting tip is for determining whether the problem is with the electronics or with the detector. A Ludlum Model 500 Pulser is invaluable at this point, because of its ability to simultaneously check high voltage, input sensitivity or threshold, and the electronics for proper counting.
Page 23
Model 12-4 Technical Manual Section 7 SYMPTOM POSSIBLE SOLUTION No power (or meter Check battery contacts. Clean them does not reach with rough sandpaper or use an engraver to clean the tips. TEST BAT OK mark) (continued) Check for loose or broken wires, especially between the main board and the high-voltage board.
Page 24
Model 12-4 Technical Manual Section 7 SYMPTOM POSSIBLE SOLUTION Meter goes full-scale Ensure that the instrument’s can is or “pegs out” properly attached. When attached (continued) properly, the speaker will be located on the left side of the instrument. If...
Model 12-4 Technical Manual Section 8 Section Technical Theory of Operation MAIN BOARD (Drawing 464 × 275 3 sheets) Input Detector pulses are coupled through C16 to emitter follower Q4. R42 provides 3.7 V bias. R41 couples to detector HV. CR3 protects the input from high-voltage transients.
Model 12-4 Technical Manual Section 8 Scale Ranging Detector pulses from the discriminator are coupled to univibrator pin 5 of U9A. For each scale, the pulse width of pin 6 of U9A is increased by a factor of 10 with the actual pulse width being controlled by the front-panel calibration controls and their related capacitors.
Model 12-4 Technical Manual Section 8 HV Supply Board (Drawing 464 × 243) Switching Convertor HV is developed by voltage multiplier CR1 through CR10 and associated capacitors. This multiplier is driven by switching convertor U2 and T1. The convertor is powered by regulated 6 V from the main board.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies –...
Model 12-4 Technical Manual Section 10 Section Parts List Reference Description Part Number Model 12-4 UNIT Completely Assembled Survey Meter Model 12 Survey Meter 48-1200 Main Board, BOARD Completely Assembled Drawing 464 × 275 Main Circuit Board 5464-275 CAPACITORS 47pF, 100V 04-5660 0.1F, 35V-T...
Model 12-4 Technical Manual Section 10 Reference Description Part Number Wiring Diagram, MTA100×5, MAIN BOARD Drawing 464 × 309 5464-275 13-8140 CONNECTORS MTA100×6, OPTIONAL 5464-275 13-8095 MTA 100×2, MAIN BOARD 5464-275 13-8073 MTA100×4, MAIN BOARD 5464-275 13-8088 MAIN BOARD 5464-275...
Need help?
Do you have a question about the 12-4 and is the answer not in the manual?
Questions and answers