Page 4
RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
MODEL 9-3 Technical Manual Table of Contents Introduction Getting Started Battery Installation Battery Test Instrument Test Operational Check Specifications Identification of Controls and Functions Safety Considerations and Maintenance Environmental Conditions for Normal Use Warning Markings and Symbols Metallized Polyester (Mylar) Window Precaution...
Page 6
Technical Manual Range Calibration Range Changing Power Supplies Audio Circuit Meter Circuit Recycling Parts List Model 9-3 Ion Chamber Main Circuit Board, Drawing 293 × 680 Electrometer Board, Drawing 293 × 670 Wiring Diagram, Drawing 293 x 525 Drawings and Diagrams...
Section 1 Section Introduction he Ludlum Model 9-3 is a five-range ion chamber instrument for measuring beta-gamma radiation up to 50 R/hr. The chamber wall, including the instrument case, is 1000 mg/cm². A 1000 mg/cm² retractable beta shield allows beta measurement with a 7 mg/cm²...
Section Getting Started Battery Installation Ensure the Model 9-3 range selector switch is in the position. Open the battery lid by pushing down and turning the quarter-turn thumbscrew counterclockwise one quarter of a turn. Install two “AA” size batteries in the compartment.
Section 2 Instrument Test Note: The Model 9-3 has two options for chamber wall voltage. An internal jumper may be selected to allow continuous voltage on the chamber wall with the instrument turned off. This option allows a faster, three-minute settling time when the instrument is frequently used.
MODEL 9-3 Technical Manual Section 2 Open the beta shield and expose the check source to the center of the chamber window. The instrument reading should be within 20% of the check source reference after settling time. Note: The depressions on the side of the instrument housing indicate the center of the chamber.
MODEL 9-3 Technical Manual Section 3 Section Specifications : typically 0.2-50,000 mR/hr. Range : reading within 10% of true value. Linearity : approximately five seconds for 90% of final meter deflection Response on the ×1 scale and ×10, and two seconds on the remaining scales Background : approximately every three minutes an up-scale “kick”...
Page 12
MODEL 9-3 Technical Manual Section 3 Warm-up Time : After power-up, ×100 through ×10K scales will be ready for use in two minutes. For the ×1 scale and if the wall voltage option is on with the instrument off, the scale will settle within 0.1 mR/hr within three minutes.
Page 13
MODEL 9-3 Technical Manual Section 3 Temperature Range: -20 to 50 °C (-4 to 122 °F) Dimensions: 23.4 x 8.9 x 21.6 cm (9.2 x 3.5 x 8.5 in.), including instrument handle Weight: 1.6 kg (3.6 lb), including batteries Ludlum Measurements, Inc.
MODEL 9-3 Technical Manual Section 4 Section Identification of Controls and Functions Range Selector Switch : This is a six-position switch marked ×1K, ×100, ×10, and ×1. Turning the range selector switch from ×10K, to one of the range multiplier positions (×10K, , ×100, ×10,...
Page 15
MODEL 9-3 Technical Manual Section 4 when the RESET switch is pressed and released. For near zero reading on ×1 scale, allow several more seconds for setting time after RESET is released Beta Shield (on bottom) : A retractable shield allowing exposure to beta radiation with a window thickness of 7 mg/cm².
The operator or responsible body is cautioned that the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by Ludlum Measurements, Inc. The Model 9-3 Ion Chamber is marked with the following symbols: The “crossed-out wheelie bin”...
MODEL 9-3 Technical Manual Section 5 Warning! The operator is strongly cautioned to take the following precautions to avoid contact with internal hazardous live parts that are accessible using a tool: 1. Turn the instrument power and remove the batteries.
Check the appropriate regulations to determine required recalibration intervals. Ludlum Measurements offers a full-service repair and calibration department. We not only repair and calibrate our own instruments, but also most other manufacturers’ instruments. Calibration procedures are available upon request for customers who choose to calibrate their own instruments.
Return for Repair or Calibration To return an instrument for repair or recalibration, provide sufficient Return Form, which can be downloaded from the Ludlum website at www.ludlums.com. Find the form by clicking the “Support” tab and selecting “Repair and Calibration” from the drop-down menu. Then choose the appropriate Repair and Calibration dvision where you will find a link to the form.
MODEL 9-3 Technical Manual Section 6 Section Technical Theory of Operation Chamber The chamber housing is constructed from acrylic and is coated inside and outside with carbon. The internal wall is maintained at approximately 90 volts. The external wall is at ground potential.
MODEL 9-3 Technical Manual Section 6 Range Change When the instrument is switched to the ×100, ×1K, or ×10K range, RL1 is closed, reducing the feedback resistance to approximately 4.2 x ohms. At 500 mR/hr on the x100 range, the chamber current will be...
MODEL 9-3 Technical Manual Section 6 Range Calibration For full-scale voltage for the electrometer output, pin 4 of U1 (electrometer board 5293-442) is approximately +0.03 volts for the ×1 range; 0.3 volts for the ×10 range; +0.025 volts for the ×100; and +0.025 volts for the ×1K range.
MODEL 9-3 Technical Manual Section 6 U4 and associated components generate +5 volts for the main board circuits. U3 provides a regulated +2.5-volt reference. VR1 provides +4.85 volts for the electrometer. U5 and associated components provides a 1.5-volt guard voltage for the electrometer.
To this end, Ludlum Measurements, Inc. strives to supply the consumer of its goods with information regarding reuse and recycling of the many different types of materials used in its products. With many different agencies –...
Technical Manual Section 8 Section Parts List Reference Description Part Number Model 9-3 Ion UNIT Completely Assembled Chamber Model 9-3 Ion Chamber 48-3633 Main Circuit Board, BOARD Completely Assembled Drawing 293 × 680 Main Circuit Board 5293-680 CAPACITORS 0.001uF, 100V 04-5659 C2-C5 0.01uF, 200V...
Page 29
4293-521 Portable Bezel with Movement Assembly 4363-188 Can Mylar Window 7293-792 Chamber Window Assembly 4293-500 Portable Handle (Rolled) 4464-154 Model 9-3 Chamber Top Co-Netic Shield 7536-223 Model 9-3 Chamber Top Netic Shield 7536-224 Ludlum Measurements, Inc. Page 8-5 June 2018...