Technical Data - Seca 763 Manual

Hide thumbs Also See for 763:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 17
the head stop slides down and does not stay in the required position?
– Place the scale carefully on its side so
that the base of the scale is accessible.
Insert the Allen key supplied in the
rectangular hole on the end of the me-
asuring rod as shown in the illustration
and tighten up the screw for the mea-
suring rod guide clockwise.
The measuring rod with the head stop
should remain freely mobile, but not be
able to slide down on its own.
Then stand the scale upright again.
9. Servicing
Your seca scale leaves the factory with an accuracy greater than ± 0,15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.

10. Technical data

Device dimensions
Width:
Height:
Depth:
Weight
Temperature range
Insulated equipment, Protection Class II:
medical electrical equipment type B
Colour
Maximum load
Graduations
Precision
Height measurement
Precision
Reproducibility
Measurement range
28
360 mm / 14.2 ins
1320 +10 mm / 52 +0.4 ins
686 mm / 27 ins
approx 20.5 kg
+10 to +40 °C
white/black
250 kg / 550 lbs / 39 sts
50 g / 0.1 lbs
better than ±0.15% / ±100g
Tolerance of ±2 mm
±1 mm
110 - 200 cm / 43 - 79 ins

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents