Error states and alarms

Error list

– Communication Board / Electronic related error messages.
Error message Tx LCD message
F116.023
Electronic Memory Failure
F108.040
Output ReadBack Failure
M030.020
Electronic Interface Error
M026.024
NV Electronic Memory Burn
Error
F106.035
Unreliable Output Current
F106.035
Unreliable Output Current
– Sensor related error messages
Error message Tx LCD message
F120.016
Sensor Invalid
F120.016
Sensor Invalid
F118.017
Sensor Memory Fail
F114.000
P-dP Sensor Fail
F112.001
Static Pressure
Sensor Fail
F110.002
Sensor Temperature
Fail
M028.018
NV Sensor Memory
Burn Error
60 OI/266/HART-EN Rev. D | 2600T Series Pressure transmitters
Possible cause
Electronic memory corrupted
The output circuit could be broken
or not correctly calibrated
Data exchange between the sensor
and the electronic is incorrect
Writings to the electronic non-Vola-
tile Memory has not been successful
The D to A converter is not properly
Calibrated/Trimmed
The Device is not properly
configured
Possible cause
The sensor signal is not being updated correctly as a
result of an electronics failure, sensor error or a poorly
connected sensor cable.
The sensor model/version is no longer compatible with the
connected electronic version
Sensor memory corrupted
Mechanical damage to the sensor. Loss of fill fluid from the
cell, ruptured diaphragm, broken sensor.
The circuitry for the sampling of the static pressure has
failed.
The circuitry for the sampling of the temperature has
failed.
Writings to the sensor non-Volatile Memory was not
successful
Suggested action
The electronic must be replaced
A DAC (digital to outpt converter) trimming
should be performed and if the error persists
the communication board must be replaced
Power off and on the transmitter and check if
the error persists. If yes replace the communi-
cation board as soon as possible.
The communication board should be replaced
as soon as possible
Perform an Output Trimming and if the error
persists the communication board must be
replaced
Check the device configuration
Suggested action
Check cable connection, check
sensor and if problem persists,
the sensor must be replaced.
The sensor must be replaced
The sensor must be replaced
The sensor must be replaced
The sensor must be replaced
The sensor must be replaced
The sensor should be replaced
as soon as possible.
Tx response
Analog Signal to Alarm
Analog Signal to Alarm
no effect
no effect
Analog Signal to Alarm
Analog Signal to Alarm
Tx response
Analog Signal to Alarm
Analog Signal to Alarm
Analog Signal to Alarm
Analog Signal to Alarm
Analog Signal to Alarm
Analog Signal to Alarm
no effect