8 Adjusting (CAL)

8.1 Adjusting for verification
General Information
Prior to the procedure of verification the balance has to be adjusted. See also section
8.2 Adjusting (CAL).
After the successful adjusting it is necessary to make the adjusting possibility
impossible on KERN EW balances. The switch for this step is located on the inside of
the balance (beneath the housing, above the display) (only with models KERN EW).
Blocking / Releasing of the adjusting function
Remark:
The adjusting is only possible when not being blocked by the adjusting switch.
Switch to the right
Switch to the left
The balance should be separated from the mains after each operation of the
switches.
After the verification the cap has to be secured by a verification mark against access.
The balance is now prepared for the procedure of verification.
EW-BA-defsi-0620
Adjusting function is blocked
After successful adjusting the adjusting has to be blocked
by this position of the switch.
This setting has to be selected before starting to verify.
Adjusting function is released